Getting Started with WaveGo

Download Instructions Here (PDF)

Instructions

Connecting for the First Time
    1. Connect the spectrometer to the phone
      Ensure On The Go cable is connected to the phone and the standard microUSB is connected to the spectrometer
    2. Connect the phone to the internet
      The phone must be connected via WiFi or a data connection in order to download the calibration file for the spectrometer
    3. Open the app
    4. Calibration files download
      The correct calibration file will automatically be downloaded from the cloud onto the phone
Take a Measurement
          1. Connect the spectrometer to the phone
            Ensure On The Go cable is connected to the phone and the standard microUSB is connected to the spectrometer
          2. Open the App
          3. Select Take New Measurement 
          4. Dark Calibration
            Follow the on-screen instructions
            Recommended every 10 measurements or if spectrum looks noisy
             
          5. Choose Measurements to calculate
            Select measurements by clicking the check boxes
            Note – further measurements can be added when reviewing the data later
          6. Take New Measurement
            Select Begin Measurement and see the data live
            To re-calibrate simply press Redo Dark
            a. Single Measurement

            (Optional) Select the number of seconds for a delay in capture
            Press Snapshot to take a measurement
            b. Continuous Measurement
            (Optional) Type the number of seconds for a delay in capture
            Press record to start the measurements
            You change the frequency and number of measurements in the settings button (Settings button at the top)

  

Save or Send a Measurement
  1. Choose "Save"
  2. Add details
    Add in optional notes and filename to identify the measurement
  3. Email results (Optional)
    Type an email address and choose whether to send raw result and/or pdf report
    Ensure there is an email app on the phone with an account signed in, and that the phone is connected to the internet
  4. Click Save
View a Saved Measurement
  1. Select View Saved Data
  2. Select Result to View or Select Multiple Measurements and Email
  3. Select Next
  4. Select View Measurement

 

Delete Saved Measurement(s) from Phone
  1. Select View Saved Data
  2. Select check box(es)
    Select the box(es) in the Delete column next to the result(s) to be deleted
  3. Select "Delete Checked"

 

Troubleshooting

Take New Measurement Button Does Not Appear

Calibration files have not been downloaded:

  • Phone must be connected to the internet before the first use to allow calibration files to be accessed

Calibration files are not available on Wave Server:

  • If using an existing spectrometer then contact Ocean Optics for details on how to use the app with an existing compatible spectrometer
Measurement Takes a Long Time

Calculating unnecessary values:

TM30 involves a lot of calculation, to speed up calculation time try unchecking the TM30 option if it’s not needed – it can be viewed from saved data if needed later

 

Light levels are low:

if the light source is not very bright then the integration time has to be longer and therefore the measurement takes longer

No Readings Appear

Measurements not chosen:

Ensure the measurements required have been selected before choosing measurement type

Connection between phone and spectrometer has been lost:

Close the app [recent apps button then swipe app from screen], disconnect and reconnect the spectrometer then try again

Emailed Results Did Not Send

No email client on phone:

There must be an email app installed on the phone with an account signed in for the email to send

Phone not connected to internet:

The phone must be connected to the internet, if not the email will be placed in the outbox of the email account until there is an internet connection

Data to send not chosen:

Ensure the check boxes for which type of data to send are selected

Email mistyped:

Ensure email address is correctly entered

Spectrum is Very Noisy

Not enough light is reaching the spectrometer:

Redo the integration time measurement from the measurement menu page

Dark calibration is incorrect:

Redo the dark measurement from the measurement menu page, ensure cap is properly in place during dark measurement

No CCT Reading

Light source is not suitable:

CCT can only be calculated for light sources that approximate black body sources, it does not work for some lights such as coloured LEDs

Calibration Load Failed

Cable connected wrong way round:

Red end - In the phone

Black end - In the WaveGo

No internet connection:

If this is your first time starting up the WaveGo, make sure you are connected to the internet so the WaveGo can access the calibration file